Calibration Procedure for the PTLevel Cistern Level Monitor

Calibration Procedure and Sensor Accuracy

After you’ve installed your new device and setup the depth of your tank, you must calibrate the device in order to achieve a proper reading. This process is simple and is explained below.

You can not calibrate at 0% or really low levels, and you must not calibrate when your device is not fully functional yet.

Calibration Procedure

First press the “Connect” button on the transmitter and refresh your browser. This obtains the latest level reading online before calibrating.

Connect Button

Measure the amount of liquid in the tank. Then divide that measurement by the full empty depth of the tank.

Example: 72″ of water / 144″ tank height = 0.5

Multiply the above result with 100 to get the current percentage.

Example: 0.5 * 100 = 50%

Enter the result into the calibration input online once. Then hit the green save icon beside the form. You can find this form under your device “Features” button on the dashboard.

Calibration

Finally, press the “Connect” button on the transmitter one more time for it to calibrate to the most recent reading.

If you have to calibrate often, you may have a different issue. Please see the other articles available to you for further troubleshooting. It is also acceptable for the reading to be off a few percent over the full range. This is due to environmental changes and the tolerances of the sensor. 

Accuracy

It is acceptable for the reading to be off a few percent over the full range. This is due to environmental changes and the tolerances of the sensor. 
The typical accuracy is ±3%

Resolution of Sensor

Pressure Range: 0 to 100kpa : Just to get an idea, 1 meter of water column is around 9.81 kpa but the tubing length and other environmental factors (such as altitude and temperature) changes that a bit.
Pressure Resolution: 0.096kpa per bit. This works to about 9.78mm of approximate resolution.

    • Related Articles

    • Calibration Procedure and Sensor Accuracy for Wired WiFi

      After you’ve installed your new device and setup the depth of your tank, you must calibrate the device in order to achieve a proper reading. This process is simple and is explained below. Jump to the end of the article for sensor accuracy and ...
    • Level Reading Maintenance - Long Range Wireless PTLevel

      Now that you have your PTLevel setup, you’re probably wondering what to do over time in order to maintain the device and it’s readings. There’s really not much to do, but as with everything, a little bit of maintenance goes a long ways. Maintenance ...
    • Monitoring small tanks

      Our 10kPa sensors have an accuracy to 0.98mm. Our regular 100kPa sensors have an accuracy to 9.78mm Custom Orders: 10kpa Sensors Pressure Range: 0 to 10kpa. The pressure resolution on a 0 to 10 kPa sensor is 0.0096 kPa per bit, which is equivalent to ...
    • Wireless Device Installation Instructions

      How to install the Long Range Wireless PTLevel device. These instructions are provided with the water level monitor that you purchase as well.
    • Temperature Sensor and the Online Account

      Let's walk through the dashboard step by step to view all of the new temperature features! Dashboard First, when you log into your account (through the App or through ptdevices.com) you'll see a summary of all of your devices. Within the summary, ...